Micro Reaction Torque Sensor 0.5 Nm to 150 Nm

Micro Reaction Torque Sensor 0.5 Nm to 150 Nm

Micro reaction torque sensor 0.5 Nm to 150 Nm

Micro reation torque sensor adopts high-quality strain gauge, aluminum alloy shell, diameter 37.6mm, minimum range 0~0.5 Nm/maximum range 0~150 Nm.

  • Whole body height 50.8mm, diameter 36.7mm
  • Shell material aluminium alloy
  • Flange and hollow shaft installation
  • 0.05%FS non-linearity, 0.05%FS hysteresis
  • Excitation voltage DC 5-15V, output signal DC 0-20mV

Specification

Model ATO-TQS-S04
Shipping weight 1kg
Capacity range 0-0.5 Nm ~ 0-150 Nm
Output signal DC 0-20mV, or equip torque sensor transmitter ATO-LCTR-OA for output signal DC 0-±5V/0-±10V/4-20mA/0-20mA
Sensitivity 2.0mV/V
Zero balance ±2%FS
Non-linearity 0.05%FS
Hysteresis 0.05%FS
Repeatability 0.05%FS
Creep 0.05%FS/30min
Temperature drift on zero <0.05%F.S/10℃
Temperature effect on output <0.05%F.S/10℃
Operating temperature -20℃~+80℃
Humidity 0~90%RH
Impedance 350Ω
Insulation resistance ≥5000MΩ/100VDC
Stability 0.3%FS/year
Safety overload 150%F.S
Ultimate overload 200%F.S
Excitation voltage DC 5-15V
Material Aluminium alloy
Protection class IP65
Cable length 2m (Φ3mm)
Wiring EXC+: Red, EXC-: Black, SIG+: Green, SIG-: White

Dimensions

Unit: mm

Micro reaction torque sensor 0.5 Nm to 150 Nm dimension

All torque sensors and repair services provided by ATO Torque Sensor are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase. Contact ATO for a complete statement of our warranty. Mechanical damage to equipment due to improper use is not covered by warranty. Electronic circuitry failure caused by the introduction of unregulated or improper excitation power or electrostatic discharge is not covered by warranty.

For proper electrical connections, refer to the installation drawing for the torque sensor, and the wiring drawing for the reaction torque sensor transmitter/amplifier used. Construct the interconnect cable from high-quality shielded instrumentation cable. Various cable assemblies are available through ATO Torque Sensor. Connect one end of the cable to the sensor connector and the other end to the sensor transmitter/amplifier. Make sure to tighten the cable connector to the sensor. For installation in dirty, humid, or rugged environments, it is suggested that the connection be shielded against dust or moisture with shrink tubing or other protective material. Strain relieving the cable/sensor connection can also prolong cable life. Mounting cables to a test structure with tape, clamps, or adhesives minimizes cable whip.

Refer to the dimension drawing supplied for specific outline dimensions and installation details.

Shunt Calibration Steps

  1. Connect the torque transducer to an appropriate torque sensor transmitter and allow adequate time for the system to stabilize.
  2. Apply a full-scale, N.I.S.T. traceable, mechanical input (or torque) to the transducer.
  3. Adjust the torque sensor transmitter’s gain or span controls, as required, to obtain a full-scale electrical output signal, and/or numeric display that represents the applied, mechanical input quantity.
  4. Remove the mechanical input (or torque).
  5. Place the shunt calibration resistor across an appropriate leg of the Wheatstone Bridge as discussed above.
  6. Record the value of the signal conditioner’s output signal and/or numeric display. This value is the shunt calibration value, or equivalent torque.
  7. It is important to note that the shunt calibration value is specific for the particular shunt resistor used. This value, and the particular resistor, are now matched to the transducer and form a basis of the transferable shunt calibration.

Failure of the sensor structure or fasteners used in its installation may cause personal injury and equipment damage. To prevent structure failure, review the extraneous load limits listed on the specification. Extraneous load limits are extraneous side force, thrust and bending moment that may be applied without electrical or mechanical damage to the torque sensor. Do not exceed moment or shear whichever attained first. Measurement inaccuracy and structure damage may result. Install the sensor in a manner that minimizes these loads. In addition, review data from manufacturers of fixtures and fasteners used in the sensors installation to determine if failure might occur due to these loads.